Unique hybrid triple quadrupole linear ion trap technology provides powerful workflows to answer complex questions with no compromise
Combination of triple quadrupole and linear ion trap analyzers in the QTRAP System enables quantitative and qualitative experiments on the same platform with no compromise. The unique configuration enables higher sensitivity and higher quality MS/MS scanning vs traditional QqQ and LIT instruments alone with expanded flexibility and unique workflows.
While there are many different types of mass spectrometers available today, the quality of the results will largely depend upon the type of mass analyzer used and the specific type of measurement or scan mode used. For targeted quantitative measurements such as multiple reaction monitoring (MRM), triple quadrupole (QqQ) instruments have long been considered the gold standard for the highest speed, sensitivity, dynamic range, and multiplexing capabilities. However, their lower duty cycle in full scan mode hinders their ability to obtain the highest performance in qualitative measurements. In contrast, QTOF or ion trap-based mass analyzers have a higher sensitivity and scan speed in full scan mode than a QqQ for qualitative measurements. The drawback however, is that these full scan-based instruments cannot typically not provide the same specifications in sensitivity, and dynamic range as a QqQ for targeted quantitative analysis.
The mass analyzer region of a QTRAP System is based on the conventional ion path of a QqQ mass spectrometer. However, in contrast to conventional QqQ systems, the third quadrupole (Q3) of a QTRAP System can also be operated as an LIT (Figure 1)1. The dual functionality of Q3 provides the QTRAP System complete functionality as a QqQ mass spectrometer but with additional powerful qualitative scan functions. These additional LIT scan functions greatly enhance the performance and flexibility for of the QTRAP System for applications such as screening, confirmation and identification applications. With a QTRAP System, all of the MRM sensitivity of a QqQ is combined with a multi-functional ion trap with no compromises. Not only does this afford the highest performance for quantitative and qualitative experiments on the same instrument, it also enables new and powerful TripleTrap™ Scanning workflows that combine QqQ functionality with LIT functionality within the same experiment.
There are two main types of quantitative scans on the QTRAP System, Multiple Reaction Monitoring and MRM3.
The triple quadrupole functionality of this instrument allows analyte quantitation to be performed at very high specificity and sensitivity using Multiple Reaction Monitoring (MRM).MRM (also known as Single Reaction Monitoring, SRM) is the scan type with the highest duty cycle and is used for monitoring one or more specific ion transition(s) at high sensitivity. Here, Q1 is set on the specific parent m/z (Q1 is not scanning), the collision energy is set to produce the optimal diagnostic charged fragment of that parent ion, and Q3 is set to the specific m/z of that fragment (Figure 2 and Figure 4, bottom). Only ions with this exact transition will be detected. Because of the speed of the LINAC® Collision Cell, many MRM scans can be looped together into one experiment to detect the presence of many specific ions in a complex mixture. Up to 4000 MRM transitions can be monitored within a single experiment and up to 500 MRMs can be monitored per second.
The area under the MRM LC peak is used to quantitate the amount of the analyte present. In a typical quantitation assay, a standard concentration curve is generated for the analyte of interest. When the unknown sample is then run under identical conditions, the concentration for the analyte in the unknown sample can be determined using the peak area and the standard concentration curve. The Scheduled MRM Algorithm Pro enables many MRM transitions to be monitored in a single acquisition through time scheduling2, with simplified, automatic method building using user supplied parameters.
There are some cases where the quantitation of specific analytes using MRM may be impaired due to co-eluting interferences or high background. Here, the use of MRM3 can greatly improve signal-to-noise, reducing the need for exotic chromatography to separate interferences or extensive sample cleanup. MRM3 is actually a hybrid QqQ/LIT workflow combining normal quadrupole functions with normal LIT functions to achieve an enhanced capability not possible with either analyzer alone3. In this scan mode (Figure 3), the precursor ion is selected in Q1 with a mass window of 1-4 amu and then fragmented in the Q2 collision cell. The fragment ions are then transferred to the ion trap where one of the resulting fragment ions is selectively isolated (isolation width of 1-5 amu). After isolation, the ion is excited using a single wavelength excitation frequency (~1 Da excitation width) to induce fragmentation (energy of excitation and excitation time are controlled by user). The resulting fragment ions are captured and then scanned out of the ion trap at one of the three scan speeds. Thus MRM3 data is collected as full scan data. Then, post-acquisition, extracted ion chromatograms (XICs) are used to isolate the secondary product ion(s) of interest for quantitation.
Two triple quadrupole scan types, the Precursor Ion Scan and the Neutral Loss Scan, can be used for identifying analytes that share a common molecular substructure. These scans allow you to survey a complex mixture for specific components and obtain the masses of these components.
When an ion loses a diagnostic fragment as a charged fragment, it can be detected by using a Precursor Ion scan. In this scan type, Q1 is scanned across a mass range, and ions are fragmented in the collision cell. Q3 is set to transmit only the mass of the diagnostic fragment ion to the detector. Therefore, only ions that are passed through Q1 that fragment to produce the diagnostic charged fragment ion will be detected (Figure 4, top).
An example of a precursor ion scan would be to monitor m/z 79 in negative ion mode in order to specifically detect phosphorylated peptides in proteomic samples. Another example would be monitoring m/z 272 in negative ion mode for the detection of glutathione (GSH) in order to detect low-level reactive metabolites in complex samples during drug discovery.
When an ion loses a diagnostic fragment as a neutral fragment, it can be detected by using a Neutral Loss Scan (NL). In this scan mode, Q1 is scanned across a specific mass range. Ions are passed into the collision cell Q2 where they are fragmented. Q3 is scanned over a similar mass range, offset by the neutral mass of the diagnostic fragment. Therefore, any molecule being passed through Q1, which loses a neutral molecule of the defined mass will then be transmitted through Q3 and detected (Figure 4, middle).
Examples of NL scans include monitoring the neutral loss of 141 to detect phosphatidylethanolamines in lipidomics experiments or monitoring a neutral loss of 176 to detect glucuronide metabolites of opiates in urine samples.
There are two full scan LIT MS scan types available on QTRAP Systems, the Enhanced MS scan (EMS), and the Enhanced Multiply Charged scan (EMC) and one narrow range high resolution MS scan, the Enhanced Resolution scan (ER). Additionally, there are two MS/MS scan types available, the Enhanced Product Ion scan (EPI) and MS/MS/MS (MS3).
The Enhanced MS scan is the standard ion trap MS scan. Ions are transmitted from the source through the quadrupoles (quadrupoles are in RF only mode) into the ion trap. After the ion trap is filled, the ions are scanned out axially to the detector. This provides a full scan analysis of all the analytes entering that QTRAP System at that point in time.
The Enhanced Multiply Charged Scan is a standard ion trap MS scan that can be used to improve the signal/noise ratio on multiply charged ions. Ions are transmitted from the source through the quadrupoles (quadrupoles are in RF only mode) into the ion trap. Once the ion trap is filled, the singly charged ions are emptied from the ion trap (~100ms), leaving predominantly the multiply charged ions behind. Ions are then scanned out axially to the detector.
The Enhanced Product Ion Scan is the standard method for performing MS/MS on the QTRAP System. It is actually a hybrid QqQ/LIT scan because it combines the capabilities of the QqQ with the LIT to achieve a performance level not possible with either analyzer alone. The fragmentation is done in the collision cell and thus provides the information rich MS/MS spectrum typical of collisionally activated dissociation (CAD) fragmentation seen on QqTOF and triple quadrupole instruments4. In this scan mode, the precursor ion to be fragmented is first selected in Q1 with a mass window of 1-4 amu wide, filtering out all other ions. The precursor ion is fragmented by CAD in the Q2 collision cell. The fragment ions generated are captured in the ion trap; then scanned out at one of three scan speeds, depending on the required fragment ion resolution (Figure 6). This provides extremely high sensitivity, high quality MS/MS data because the fragment ions are captured and analyzed in the LIT. However, it is important to note that because the mass isolation step is performed in Q1 and not within the ion trap, a complete spectrum is obtained covering the full mass range. This is in contrast to when isolation and fragmentation are performed within the ion trap itself where masses from the lower 1/3 of the parent ion will be “cut off” and unobservable due to the energetics required for the experiment. Thus an EPI scan does not suffer from the low mass cut-off problem typical of other ion trap mass analyzers (Figure 7).
The MS/MS/MS Scan is a very powerful tool for obtaining further structural information on a compound of interest. This scan is in practice performed similar to the MRM3 scan. Q1 selects a precursor of interest. Q2 fragments the precursor by CAD. Fragment ions are passed to the LIT where one is selectively isolated to induce further fragmentation. The resulting secondary fragment ions are then scanned to the detector. The full scan collected on the secondary fragment ion can be used to further distinguish the structure of the target analyte.
The unique configuration of the QTRAP Systems also enables powerful workflows that combine quantitative and qualitative capabilities within the same experiment. Here QqQ and LIT scans can be linked in a cyclic manner to provide unmatched power and flexibility for a multitude of applications.
TripleTrap Scanning refers to any workflow that engages both a QqQ scan and LIT scan within the same experiment. Typically TripleTrap Scanning is employed during information dependent acquisition (IDA) experiments in which a QqQ scan is followed by an LIT scan. Some examples of TripleTrap Scanning workflows include:
Q0 trapping can be employed to increase the sensitivity of any LIT or hybrid scan. When Q0 trapping is turned on, ions produced at the source are accumulated in Q0 during the scanning of the LIT. When the ion trap is ready to be filled again, the ions from Q0 are released (Figure 8). Often, ~5-10x increase in signal intensity is observed in MS/MS mode when Q0 trapping is activated.
Dynamic Fill Time (DFT) is a feature specifically designed to optimize the data obtained in every spectrum for the linear ion trap scans. DFT will automatically adjust the fill time used to fill the ion trap based on the ion flux coming from the source. A very quick quadrupole scan is performed using the same mass range to be measured and determines the number of ions reaching the detector. For more intense ions, the fill time will be automatically reduced to ensure the trap is not overfilled with ions. For less intense ions, the fill time will be automatically increased, ensuring that good ion statistics are obtained in the spectrum.
Collision Energy Spread is a feature designed to improve the fragmentation quality of the MS/MS data collected ‘on the fly’ during Information Dependent Acquisition (IDA). The collision energy used during the filling of the ion trap is stepped across a range of energies preset in the software, ensuring that an information rich spectrum is obtained during every MS/MS.
The unique combination of a QqQ and LIT analyzer within the QTRAP System enables both quantitative and qualitative experiments on the same platform with the same performance one would expect from two separate high quality, high performance instruments. High sensitivity quantitation experiments can be performed in MRM mode with no compromise in performance. When required, the unique configuration can be leveraged to obtain qualitative information on samples. In fact, the unique configuration of the two analyzers together actually enables higher sensitivity and higher quality MS/MS scanning beyond what is possible with traditional triple quadrupole and linear ion trap instruments alone. Additionally, the QTRAP System configuration enables valuable new workflows that can link QqQ scans with LIT scans within automated acquisition strategies to provide the ultimate flexibility for a wealth of different analytes and applications.