Using SCIEX QTRAP systems and technology
Liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS) is a widely used analytical tool for screening and confirmation of drugs of abuse in forensic samples. The detection of several compound classes, such as the common pain drug panel and drugs of abuse is very important. While this analysis has generally been performed utilizing urine sample matrices, many research laboratories are investigating the analysis of these compounds in oral fluid matrices. Using oral fluid as a matrix provides several advantages over urine analysis. It provides a simpler collection procedure that can be witnessed which means there is less opportunity to adulterate the sample. Typical collection of oral fluid matrix involves a collection swab that retains the oral fluid matrix. After collection, the swab is subjected to extraction buffer. The resulting matrix is then analyzed for the drug compounds. The outcome for these investigations into the use of oral fluid needs to provide a means to accurately and precisely quantify compounds in a timely fashion, while having a simplified workflow that is conducive to their setting. This current effort has shown to accomplish just that. In this study we present a fast, robust, and reliable method that has afforded the detection of 41 compounds in an oral fluid matrix. The compounds included 6-MAM, alprazolam, amphetamine, benzoylecgonine, buprenorphine, carisoprodol, clonazepam, codeine, diazepam, EDDP, fentanyl, flunitrazepam, flurazepam, hydrocodone, hydromorphone, hydroxyalprazolam, lorazepam, MDA, MDEA, MDMA, meperidine, meprobamate, methadone, methamphetamine, midazolam, morphine, naloxone, naltrexone, norbuprenorphine, nordiazepam, norfentanyl, normeperidine, norpropoxyphene, oxazepam, oxycodone, oxymorphone, PCP, propoxyphene, sufentanil, temazepam and tramadol.
Calibrator preparation: An oral fluid matrix was spiked to various levels of the relevant compounds. The calibration curve extended above and below the industry accepted cutoff and confirmation levels.
Sample preparation:
HPLC conditions: Chromatography was performed using a Phenomenex Kinetex C18 2.6 µm 50 x 3.0 mm column using ammonium formate in water as mobile phase A and formic acid in methanol as mobile phase B. A gradient was applied during chromatographic run and the total run time (not multiplexed) per stream was 6 minutes. The acquisition window was set to 3.5 minutes per stream (Figure 2) providing an increase in sample throughput of 171% during multiplexed analysis.
MS/MS conditions: MS/MS analysis was carried out using a QTRAP System operating in positive electrospray ionization mode. Analytes were analyzed using the Scheduled MRM Algorithm to accommodate for quantification of all components listed above.1 In addition to the Scheduled MRM Algorithm (sMRM) quantification, an information dependent acquisition (IDA) method was set up for targeted drug screening which involved an sMRM survey scan followed by enhanced product ion scans (EPI) for collection of full scan MS/MS for identification.
Figures 1 and 4 show typical chromatograms achieved using the described approach and quantitative results for the following analytes: codeine, methamphetamine, midazolam and temazepam. Excellent linearity and reproducibility were observed throughout the dynamic range assessed in this effort. Table 1 shows a summary of the calibration range for all 41 pain drugs. Within the listed concentration range, excellent linearity and reproducibility were achieved.
Figure 5 shows an excerpt from a report generated for the results of the analysis performed using a QTRAP LC-MS/MS System in which both the triple quadrupole and linear ion trap functionalities have been combined in one MS method to perform both quantification of detected compounds and confident identification of the compounds through MS/MS spectral library matching.
The QTRAP LC-MS/MS System combined with the MPX 2.0 High Throughput System is a powerful tool for ultra-fast multi-analyte quantification and identification in forensic samples. A fast, robust and simple method for the detection of 41 common pain drugs in an oral fluid matrix was developed on this system, providing the high throughput required for today’s modern laboratories. Using the Scheduled MRM Algorithm in combination with acquisition of full scan MS/MS spectra for compound identification through MS/MS library spectral searching enables large numbers of analytes to be quantified in a single method.