High speed acquisition with TripleTOF® 6600 System
Nick Morrice1, Zuzana Demianova2, Xiaolei Lv3, Christie Hunter4
1SCIEX, UK, 2SCIEX, Germany, 3SCIEX, China, 4SCIEX, USA
Increasingly, microflow liquid chromatography (LC) is being used for the analysis of complex proteomics samples because of higher robustness and faster run times. Here, the use of very fast microflow LC gradients was explored for protein identification experiments, using gradients as short as five minutes. Combined with MS/MS acquisition rates as fast as 66 Hz on the TripleTOF® 6600 System, good protein identification results are achieved providing high sample throughput. Using multiple instruments and multiple complex matrices, it was observed that the number of proteins identified for even the fastest gradients was still over 1000 for multiple complex matrices from 1 µg protein load.
One of the primary challenges to producing statistically significant results for proteomics has been the long times required for data acquisition using nanoflow chromatography. Even relatively simple protein identification experiments require up to one hour of instrument time per sample due to long sample injection, column washing and equilibration times. Microflow liquid chromatography (LC) provides more reproducible separations and more robust electrospray, and is being used increasingly for the analysis of complex proteomics samples. Additionally with higher flow rates, less time is required to load, wash and equilibrate the trap and column.
Data presented here demonstrates the high-throughput workflows now possible for data dependent acquisition (DDA) using microflow liquid chromatography gradients as short as five minutes (total run time <15mins). The combination of microflow LC with the high MS/MS acquisition rates of the TripleTOF® 6600 System enables high-throughput yet comprehensive analysis for proteomics samples, at rates approaching 100 samples per day.
Using multiple instruments and multiple complex matrices, it was observed that the number of proteins identified for even the fastest gradients was still over 1000 for multiple complex matrices from 1 µg protein load (Figure 1).
Sample preparation: Cell lysates (HEK, K562 and Yeast) were digested with trypsin using standard protocols. Sample loading of 0.5 – 2 µg of total protein were used for each injection.
Chromatography: A NanoLC™ 425 System plumbed for microflow chromatography was used (5 µL/min) and operated in trap/elute mode. Column temperature was controlled at 30°C. Gradients of 5, 10, 20, or 45 minutes were tested (Table 1). More information on LC configuration can be found in the SWATH Performance Kit SOP.1
Mass spectrometry: All data was acquired using a TripleTOF® 6600 System with the Turbo V™ Source equipped with the 25 µm hybrid electrodes for microflow LC. IDA data were collected using a variety of acquisition strategies. The TOF MS scan was 150 msec and the number of MS/MS per cycle and accumulation time was varied. Other method parameters were varied as described for Figure 3.
Data processing: Data dependent acquisition (DDA) data was processed using ProteinPilot Software 5.04 and the FDR reports were evaluated using the FDR Comparison Template2 provided with the software.
A series of optimization experiments were performed to test the best acquisition conditions for running data dependent experiments under accelerated chromatographic conditions. As faster gradients will produce taller sharper LC peaks, parameters such as intensity threshold for MS/MS trigger, number of MS/MS per cycle and accumulation time (total cycle time), exclusion time after MS/MS would need to be adjusted.
Figure 3 shows a few examples of the parameter optimization ramps that were performed. Increasing the number of MS/MS per cycle provided significant improvements in the number of IDs. The accumulation time per MS/MS was also decreased to maintain cycle time and it was found that very fast acquisition rates (accumulation times down to 15 msec) provided results improvements. There was little dependence on intensity threshold, perhaps due to the overwhelming # of precursors available in every cycle, however the threshold was increased to 500 or 1000 for the very fast MS/MS experiments to ensure data was acquired at points higher up on the LC peak. At these very fast accumulation times (15 msec), ~70% of the acquired spectra yielded peptide identifications, indicating the high spectral quality.
Optimized collision energy values for peptides3 were used, and for the very low accumulation times, collision energy spread (CES – short CE ramp around optimized values) was turned off. This was found to have minimal impact on results quality.
A few different protein loads were tested using these faster gradients to understand the impact on protein identification rates. Figure 4 shows the results from the 10 and 20 minute gradients across 3 different protein loadings for K562 digest on one of the TripleTOF Systems tested. The number of peptides identified increased by about 13% when the load was increased from 0.5 to 2 µg resulting in an 8% increase in protein IDs.
Using a protein load of 1 µg, the acquisition conditions determined best for each gradient length, a final dataset was collected to show the ID trends (Figure 1). Here, the numbers shown were generated using the acquisition method with 90 MS/MS per cycle and 15 msec accumulation time, providing 66 Hz acquisition rates for the 5, 10 and 20 min gradient. 60 MS/MS x 25msec was used for the 45 minute gradient. The number of proteins and peptides identified by ProteinPilot Software at <1% global FDR5 were reported.
Very similar numbers of peptides were detected between the 3 instruments on the different matrices (Figure 5). Very little decrease is seen as the gradient was shortened from 45 to 20 mins. At 10 mins, about 9000 peptides are observed at < 1% global FDR. Even for the 5 min gradient, 5000 peptides were identified.
The relative numbers of proteins identified at the various gradients lengths were plotted as well to provide researchers a guidance when selecting the right LC strategy for a study (Figure 6). Comparing to the protein numbers obtained with the typical 45 min gradient used for microflow, small decreases in protein numbers were observed at shorter run times. However, even with the shortest gradients of 5 mins, 50% of proteins were still identified, suggesting this is a compelling strategy for fast interrogation of proteomics samples.
Microflow LC is proving to be a robust and useful tool for advancing and accelerating proteomics research. Here the use of very fast LC gradients was explored, using microflow LC gradients as short as five minutes. Combined with MS/MS acquisition rates as fast as 66 Hz on the TripleTOF 6600 System, good protein identification results are achieved providing high sample throughput.
Protein ID rates were nearly as good for 20 min as for a 45 min gradient for both peptides and proteins using microflow LC. ID rates drop using faster gradients but are still quite high when the 10 and 5 minute gradients are used (ie. for 10 min, ID ~50 and 75% respectively of peptides and proteins found at 45 mins).
When more rapid protein ID results are required or when less complex samples are to be analyzed, the fast microflow LC approach with TripleTOF Systems provides a high-throughput robust solution.